Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.7 Manual reprocessing of the water-conducting system
2.7 Manual reprocessing of the water-conducting
system
Note
If the "Water block compact", a water bottle for the Water block DVGW as-
sembly kit with integrated water disinfection unit is used without the intens-
ive germ reduction assembly kit, the instrument rinsing programmes 01 and
02 can be carried out.
Risk of infection if the "Water block, compact" is used without addi-
tional safeguards.
Contamination of the treatment water and/or drinking water supply with
germs.
▶ With regard to the "Water block, compact" assembly kit, please note that
no germ reduction facility is installed in the unit, and take appropriate safe-
guards. KaVo recommends to use the "Water block DVGW with integrated
water germ reduction system" in combination with KaVo OXYGENAL 6
(Mat. no. 0.489.3451).
▶ If the Water bottle kit is used with the enclosed dosing attachment (Mat.
no. 1.002.0287), add the proper amount of KaVo OXYGENAL 6 (Mat. no.
0.489.3451) with each filling. For the correct amount, please refer to the
instructions of the dosing attachment for water germ reduction.
Manual reprocessing procedure:
▪ Before first start-up and after the device has not been used for a while
(weekends, holidays, vacation times, etc.)
▪ Before you start working
▪ After each patient (short rinsing programme of 20 seconds)
▶ Rinse all water outlets without handpieces for 2 minutes and actuate tum-
bler filler repeatedly.
▶ Before first start-up and after the device has not been used for a while,
briefly blow air through the air lines.
2.8 Automatic hygiene functions of the water
conducting system
Note
Intensive germ reduction kit and water block DVGW required.
Health damage due to germ formation.
Infection hazard.
▶ Before starting, rinse all the water drain lines without instruments.
▶ Before start-up and after the device has not been used for a while (week-
ends, holidays, vacations, etc.), rinse or purge with air the air and water
lines.
▶ Carry out an intensive germ reduction.
▶ Actuate the tumbler filler repeatedly.