2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit

2.6.3 Preparation of the spittoon

Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶ Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
▶ Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.
Cleaning and disinfection of the spittoon bowl
Damage caused by inappropriate cleaning agents.
▶ Do not use acid-containing, sand-containing, and form-forming cleaning
agents.
▶ Add a dose of DEKASEPTOL gel to the spittoon bowl and distribute the
agent with a brush.
▶ For cleaning, allow the gel to act for approx. 5 minutes.
▶ For disinfection, allow the gel to act for approx. 15 minutes.
▶ After letting it work, press the "Spittoon bowl" button.
▶ Clean the spittoon bowl thoroughly using a brush.
▶ For thorough cleaning or thermodisinfection, rotate the spittoon bowl and
lift it off.
Damage caused by treatment in a washer disinfector.
▶ Do not subject spittoon bowls made of glass to washer disinfector treat-
ment.
The spittoon bowl made of porcelain can be subjected to thermodisinfection.
NOTICE
NOTICE
NOTICE
23 / 44
Care instructions Primus 1058 Life