2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys-
Place the COMFORTbase on the rinsing attachment
Carry out the rinsing programme
Set rinsing time
M
T
With the foot pedal of the foot control being pressed down, the rinsing time for
the respective rinsing programme can be set in the range from 00:01 to 4:00
minutes using the "Decrease value" and "Increase value" buttons.
▶ Insert the triple-function or multifunctional handpieces without the cannula
into the clamping device of the germ reduction attachment. Make sure that
the "W" button (water) of the triple-function or multifunctional handpiece is
activated, and that no air is exiting.
▶ Place the instrument hoses of the dentist and assistant elements in the cor-
responding places of the germ reduction attachment.
▶ Support tubing with the enclosed holder.
Note
All water regulation mechanisms for the motor, instruments, and multiflex
couplings must be open.
Note
A seat adapter for the germ reduction attachment is available for the COM-
FORTbase.
▶ For the rinsing programme and intensive germ reduction, press the seat ad-
apter into one of the four rubber seats of the germ reduction attachment.
▶ Place the COMFORTbase in the seat adapter.
▶ Press the "decrease value" or "increase value" keys to select rinsing pro-
gram 01 "Programm vorbereiten: 01" or rinsing program 02 "Programm
vorbereiten: 02".
ð
The standard value for rinsing program 01 is 00:20 minutes per instrument.
ð
The standard value for rinsing program 02 is 2:00 minutes per instrument.
▶ Press the foot pedal to display the time for rinsing the instrument hoses and
the three-function or multifunctional handpiece.