2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys-
M
T
Intensive germ reduction during extended periods of
non-use (> 4 weeks)
During extended periods of non-use, e.g. semester breaks at university, intens-
ive germ reduction must be carried out during the time of non-use. The germ
reduction fluid remains inside the system during the period of non-use.
Before the period of non-use
After the period of non-use
Additional intensive germ reduction during period of
non-use
KaVo recommends carrying out an additional intensive germ reduction at uni-
versities during semester breaks. This flushes the parts of the unit and replaces
the germ reduction liquid.
▶ Simultaneously press the foot pedal on the foot control.
ð
The intensive germ reduction switches to step 3.7.
ð
The automatic rinsing process starts up.
▶ Start the intensive germ reduction and stop it in step 4 after the signal tone
during the dwell time of 30 minutes.
▶ To prevent damage due to hoses getting kinked, remove hoses from the
germ reduction attachment and place them in the dentist support.
▶ Keep the germ reduction attachment in the spittoon bowl.
▶ Cover the treatment unit during extended periods of non-use.
▶ Turn on the central compressor.
▶ Switch on the water supply for the treatment unit after the system pressure
is achieved.
▶ Replace the instrument hoses of the dentist and assistant units in the places
provided in the germ reduction attachment.
▶ Turn the treatment unit on.
ð
The unit automatically starts the remaining steps of the unfinished cycle.
▶ Allow the cycle to complete and then carry out a second full intensive germ
reduction.
ð
The unit is now ready for use.
31 / 44
Care instructions Primus 1058 Life