Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment
units
At the site of use
Storage and transport
Preparations for cleaning
Manual cleaning
Automatic cleaning
Manual disinfection
Automatic disinfection
2.2 General instructions for the reprocessing of KaVo
treatment units
Please comply with the general reprocessing instructions at all times. Detailed
product-specific reprocessing instructions are provided subsequently.
▶ Remove any surface contamination using a disposable cloth/pa-
per towel.
KaVo recommends reprocessing as soon as possible after use.
▶ Comply with product-specific reprocessing instructions.
Fixtures:
▪ Brush or cloth
▪ Tap water 30 °C ± 5 °C
▶ Remove any dirt on the surface using a suitable cleanser accord-
ing to the recommendations of the manufacturer. Before doing
so, check the compatibility of the cleanser on a inconspicuous
place of the treatment centre.
KaVo recommends washer disinfectors from Miele in accordance
with EN ISO 15883-1 that are operated with alkaline cleaning
agents.
The validations were conducted with the VARIO-TD programme, the
cleaning agent neodisher® MediClean and the neutralisation agent
neodisher® Z.
▶ For programme settings as well as cleansers and disinfectants to
be used, please refer to the Instructions for Use of the washer
disinfector.
▶ Directly after automated cleaning/disinfection, treat the medical
device with the care products and systems provided by KaVo.
▶ Perform wipe disinfection only, the use of spray disinfection and
disinfection baths is not feasible.
▶ Wipe all outer surfaces with a soft cloth and an approved disin-
fectant.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ FD 322 (Dürr Dental)
▪ Incidin Liquid (Ecolab)
▶ Use disinfectants according to the manufacturer's Instructions for
Use.
▶ Please note the disinfectant safety data sheet.
KaVo recommends washer disinfectors from Miele in accordance
with EN ISO 15883-1 that are operated with alkaline cleaning
agents.
The validations were conducted with the VARIO-TD programme, the
cleaning agent neodisher® MediClean and the neutralisation agent
neodisher® Z.
▶ For programme settings as well as cleansers and disinfectants to
be used, please refer to the Instructions for Use of the washer
disinfector.
▶ In order to prevent negative effects on the medical device, make
sure that the interior and the exterior of the medical device are
dry after completion of the cycle, and then grease it immediately
with servicing agents from the KaVo care system.
6 / 44