Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
Disinfection of the 3-function and multifunction
handpiece
Manual disinfection
External disinfection
▶ Wipe-disinfect the surfaces with a soft cloth.
Automatic disinfection
Not applicable.
Damage caused by improper cleaning/disinfection.
▶ Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶ Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶ Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.
Drying the 3-function and multifunction handpiece
Manual drying
▶ Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic drying
Not applicable.
Care of the 3-function and multifunctional handpiece
Requirement
The multifunctional handpiece has been reprocessed.
▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2
(Mat. no. 0.411.7510) to the O-rings of the joint between the gripping
sleeve and cannula. Use cotton buds for this purpose.
Sterilising the 3-function and multifunction handpiece
Note
The grip sleeve and cannula can be sterilised.
▶ Hold the 3-way or multifunctional handpiece at the gripping sleeve and take
off the cannula with a slight twisting motion.
NOTICE
16 / 44