Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit
Cleaning and disinfection of the tumbler holder and
filler
Note
When assembling, make sure that the O-ring is present on the tumbler filler.
Note
Inspect the O-ring (Mat. no. 0.200.6128) for damage and exchange if ne-
cessary to prevent water or oxidation damage.
▶ Remove the tumbler filler ① (Mat. no. 0.764.8541) by pulling upwards.
▶ Take off the tumbler holder ③ (Mat. no. 07646153).
▶ Clean the tumbler holder and tumbler filler under running water.
▶ Wet the O-ring ② (Mat. no. 02006128) with silicone grease (Mat. no.
10006403).
Damage caused by treatment in a washer disinfector.
▶ Do not subject the instrument holder to washer disinfector treatment.
The tumbler filler can be subjected to thermodisinfection.
Sterilisation of the tumbler holder and filler
Damage caused by sterilisation.
▶ Do not sterilise the tumbler holder and filler.
NOTICE
NOTICE
22 / 44