2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
▶ Clean the surface with a soft cloth and mild liquid cleaning agent.
Disinfection
Exterior disinfection by hand
Highly flammable gases-.
Explosion.
▶ Allow product to cool before disinfecting.
Damage to the camera due to improper disinfection.
Irreversible destruction of the camera.
▶ Disinfect with wipes only, do not use disinfectant sprays!
▶ Do not immerse the camera in liquids!
▶ Do not sterilise the camera in an autoclave!
▶ Do not store the camera in Steri pack!
▶ Do not sterilise the camera with hot air or in a washer disinfector!
▶ Switch main device switch off.
▶ Allow the camera to cool down.
▶ Remove the camera.
▶ Disinfect the surface using a soft wipe and disinfectant.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
WARNING
NOTICE
19 / 44
Care instructions Primus 1058 Life