Instructions for use
MASTERmatic LUX M10 L - 1.009.3570

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MASTERmatic LUX M10 L and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for KaVo MASTERmatic LUX M10 L

  • Page 1 Instructions for use MASTERmatic LUX M10 L - 1.009.3570...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............8 2.5 Qualification of personnel....................
  • Page 4: User Instructions

    MASTERmatic is a registered trademark of KaVo Dental GmbH. All other trademarks are property of their respective owners. KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.kavobox.com. KaVo Technical Service...
  • Page 5 Instructions for use MASTERmatic LUX M10 L - 1.009.3570 1 User instructions Serial number UDI symbol Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity)
  • Page 6 Instructions for use MASTERmatic LUX M10 L - 1.009.3570 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. CAUTION In cases which – if not prevented – can lead to material damage.
  • Page 7: Safety

    4 Do not operate the medical device at eye level. 2.3 Technical condition A damaged product or damaged or NOT KaVo original components could injure patients, users or third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
  • Page 8: Accessories And Combination With Other Equipment

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 2 Safety | 2.4 Accessories and combination with other equipment ▪ Malfunctions ▪ Damage (e.g., from dropping) ▪ Irregular running noise ▪ Excessive vibration ▪ Overheating ▪ Dental bur is not seated firmly in the handpiece...
  • Page 9: Qualification Of Personnel

    Define the service interval depending on the frequency of use. As a result of the use of NON-KaVo original spare parts during the repair, parts may become detached and injure patients, users or third parties. This may res- ult in aspiration, swallowing of parts and possibly even a risk of suffocation.
  • Page 10: Description Of The Product

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 3 Description of the product | 3.1 Intended use 3 Description of the product MASTERmatic LUX M10 L (Mat. no. 1.009.3570) 3.1 Intended use Indications for use: This medical device is: ▪...
  • Page 11: Technical Specifications

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 3 Description of the product | 3.2 Technical Specifications 3.2 Technical Specifications Drive speed max. 40,000 rpm Speed transmission 1 : 1 Identification 1 blue ring Spray water pressure 0.8 to 2.0 bar (12 to 29 psi) Spray air pressure 1.0 to 2.0 bar (15 to 29 psi)
  • Page 12: Startup And Shut Down

    Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management / recycling companies. Comply with the comprehensive return system. KaVo has had its packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
  • Page 13 Instructions for use MASTERmatic LUX M10 L - 1.009.3570 4 Startup and shut down | 4.1 Checking the amount of water 13 / 28...
  • Page 14: Operation

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental bur. Cuts, infection and burn injury. 4 Do not touch the dental bur while it is rotating. 4 Remove the dental bur from the handpiece after treatment to avoid injury and infection during storage.
  • Page 15: Inserting The Dental Bur

    CAUTION Dental bur with damaged, worn or deformed shafts. A damaged product or damaged or NOT KaVo original components can injure patients, users or third parties. Injury hazard, dental bur may fall out during treatment. 4 Never use dental burs with damaged, worn or deformed shafts.
  • Page 16: Removing The Dental Bur

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 5 Operation | 5.4 Removing the dental bur CAUTION Dental bur with damaged, worn or deformed shafts. Material damage to the chuck system, dental bur is difficult or impossible to re- move from the chuck system.
  • Page 17: Checking For Malfunctions And Troubleshooting

    Malfunction or complete failure. 4 Do not use Vaseline or other grease or oil. 4 Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the coupling. 4 Press the O-ring between your fingers to form a loop.
  • Page 18: Cleaning The Spray Nozzle

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses. 6.2.2 Cleaning the spray nozzle CAUTION Hazard from insufficient amount of spray water.
  • Page 19: Reprocessing Steps In Accordance With Iso 17664-1 / Iso 17664-2

    Malfunction and material damage. 4 Reprocess in a washer disinfector or by hand only. KaVo recommends the following products based on the compatibility of the ma- terials. The microbiological efficacy must be ensured by the disinfectant manu- facturer and proven by an expert opinion.
  • Page 20: Manual Reprocessing

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual Reprocessing 7.2 Manual Reprocessing WARNING Incomplete disinfection. Patients, users or third parties could be infected by contaminated medical devices.
  • Page 21: Manual Pre-Cleaning

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual Reprocessing 7.2.1 Manual pre-cleaning 4 Rinse the instrument for 20 seconds with process water on the treatment centre.
  • Page 22: Manual Drying

    7.2.3 Manual drying Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Hold the can vertically.
  • Page 23: Automated Drying

    NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 4 Remove any residual liquids with KaVoDRYspray.
  • Page 24: Packaging

    Also refer to: Instructions for use KaVo QUATTROcare PLUS Servicing the chucking system KaVo recommends servicing the chuck system once weekly using the chuck servicing programme integrated in the device. NOTE Handpieces must be taken off the service couplings before the chuck service can be started and performed.
  • Page 25: Storage

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.7 Storage CAUTION Contact corrosion due to moisture. Damage to product. 4 Immediately remove the product from the steam steriliser after the steril- isation cycle.
  • Page 26: Optional Aids And Consumables

    Instructions for use MASTERmatic LUX M10 L - 1.009.3570 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862...
  • Page 27: Terms And Conditions Of Warranty

    24 months from the date of the invoice, subject to the following conditions: With regard to justified complaints KaVo grants warranty in the form of a free of charge repair or delivery of a replacement. Other claims of any kind whatso- ever, in particular with respect to compensation, are excluded.