Instructions for use SURGmatic S11 L – 1.009.1010, SURGmatic S11 C – 1.009.1005
4 Startup and shut-down | 4.1 Checking the water quantity

4 Startup and shut-down

Hazard from contaminated products.
Patients, users or third parties can be infected by contaminated medical de-
vices.
4 Prior to initial startup and after each use, process the product and acces-
sories.
Dispose of the product in the appropriate manner.
Patients, users or third parties can be infected by contaminated medical de-
vices.
4 Process the product and accessories before disposal.
Also refer to:
7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2, Page 20
Current packaging law
Dispose of the packaging properly in accordance with the current packaging law
using disposal companies/recycling firms. Comply with the comprehensive re-
turn system. KaVo has had its packaging licensed for this purpose. Please com-
ply with the regional public waste-disposal system.

4.1 Checking the water quantity

Overheating of the tooth due to insufficient amount of cooling water.
Insufficient spray water can cause the medical device to overheat and damage
the pulp and tooth.
4 Adjust the water amount for the spray cooling to a minimum of 50 ml/min
3
(3.1 inch
).
4 Check spray water channels and, if applicable, clean spray tubes with the
nozzle needle (Mat. No. 0.410.0931).
4 Ensure that the coolant supply is free of air.