Preparation Of The Dentist Unit; Reprocessing Of Handle And Holder Pad/Anti-Slip Mat - KaVo Primus 1058 Life Care Instructions

Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
Damage caused by spray disinfection.
Disinfectant gets into cracks and loosens the paint.
▶ Do not use spray disinfection.
▶ Only disinfect by wiping.
Cleaning and disinfection of the foot control
Damage caused by inappropriate use of water.
Damage to the foot control.
▶ For cleaning, the foot control must not be immersed in water or placed un-
der running water.
▶ When wiping it off with a moist cloth, do not touch the contacts of the
charge socket.
▶ Only wipe off. Do not spray.

2.4 Preparation of the dentist unit

2.4.1 Reprocessing of handle and holder pad/anti-slip
mat
Note
The handle of the TM table cannot be detached.
Cleaning and disinfection of the handle TM table / cart
▶ Clean the handle with a soft cloth and water, and wipe-disinfect it with an
approved surface disinfectant.
NOTICE
NOTICE
9 / 44