Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys-
Place the instruments on the sterilisation attachment
with patient part assembly kit / without patient part assembly kit
The cannula of the triple-function and multifunctional handpieces must be re-
moved for the rinsing program and the intensive germ reduction.
S-table
▶ Lock the swing-arm into the front catching position to ensure that the hand-
piece is removed.
▶ In the case of swing-arms without a catching function, insert the swing-arm
lock (Mat. no. 1.008.1511) to ensure that the handpiece is removed.
▶ Hold the 3-way or multifunctional handpiece at the gripping sleeve and take
off the cannula with a slight twisting motion.
26 / 44