Table 3-10 Digital code and voltage output table for bipolar setting (U2351A and U2353A)
Digital Code (Hex)
0xFFFF
0x8001
0x8000
0x7FFF
0x0000
Table 3-11 Digital code and voltage output table for unipolar setting (U2351A and U2353A)
Digital Code (Hex)
0xFFFF
0x8000
0x0001
0x0000
U2300A Series USB DAQ User's Guide
Analog Output
Vref * (32767/32768)
Vref * (1/32768)
0 V
–Vref * (1/32768)
–Vref
Analog Output
Vref * (65535/65536)
Vref * (32768/65536)
Vref * (1/65536)
Vref * (0/65536)
Voltage output (with internal reference of +10 V)
9.999694 V
0.000305 V
0 V
–0.000305 V
–10.000 V
Voltage output (with internal reference of +10 V)
9.999847 V
5.00000 V
0.000152 V
0 V