3
Features and Functions
50
16-bit AI range
The following
Table 3- 5
transfer characteristics of bipolar and unipolar input ranges
of the U2351A, U2352A, U2353A, U2354A, U2355A, and
U2356A models.
Table 3-5 Analog input range and digital code output for bipolar
Description
Full-scale Range (FSR)
±10 V
Least significant bit
305.2 µV
(LSB)
FSR–1LSB
9.999695
Midscale+1LSB
305.2 µV
Midscale
Midscale–1LSB
–305.2
–FSR
–10 V
Table 3-6 Analog input range and digital code output for unipolar
Description
Full-scale Range
0 V to 10 V 0 V to +5 V 0 V to +2.5 V 0 V to +1.25 V
(FSR)
Least significant
152.6 µV
bit (LSB)
FSR –1LSB
9.999847 V 4.999924 V
Midscale +1LSB
5.000153 V 2.500076 V
Midscale
5 V
Midscale –1LSB
4.999847 V 2.499924 V
–FSR
0 V
and
Table 3- 6
Bipolar analog input range
±5 V
±2.5 V
152.6 µV
76.3 µV
4.999847
2.499924
V
V
V
152.6 µV
76.3 µV
0 V
0 V
0 V
–152.6
–76.3 µV
µV
µV
–5 V
–2.5 V
Unipolar analog input range
76.3 µV
38.15 µV
2.499962 V
1.250038 V
2.5 V
1.25 V
1.249962 V
0 V
0 V
U2300A Series USB DAQ User's Guide
describes the ideal
Digital code
output
±1.25 V
38.15 µV
1.249962
7FFF
V
38.15 µV
0001
0 V
0000
–38.15
FFFF
µV
–1.25 V
8000
Digital code
output
19.07 µV
1.249981 V
7FFF
0.625019 V
0001
0.625 V
0000
0.624981 V
FFFF
0 V
8000