Check 10, Dc-Dc Converter Input Voltage - Agilent Technologies 4338B Service Manual

Milliohmmeter
Hide thumbs Also See for 4338B:
Table of Contents

Advertisement

Check 10, DC-DC Converter Input Voltage

1. Turn on the 4338B.
2. Using a multimeter, check if the voltage between \DC OUT" test pad and TP500 (DC-DC
Converter input voltage) is between 12 V and 15 Vdc. Refer to Figure 4-13 for test pad and
pin locations.
3. If the voltage is in range, proceed to the \Check 11, DC-DC Converter Open Output
Voltages", otherwise proceed to the \Check 12, Transformer Secondary Voltage".
T roubleshooting
4-17

Advertisement

Table of Contents
loading

Table of Contents