AMETEK CTS
12.4.4 Immunity from Continuous Power Line Disturbances: CI 210
The device shall be immune from continuous disturbances that occur on the vehicle's low voltage (i.e. 13.5 VDC)
electrical distribution system.
Default test setup
RI 210 Test Parameter requirements
Range
Frequency [kHz]
0.05 – 1
Range 1
1 – 10
Range 2
Range 3
10 - 100
Manual for Operation
Figure 19 Setup for CI 210 closed loop with scope
Figure 20 Graph for CI 210
Us [Vp-p]
Frequency step [kHz]
0.2
0.2 x f
2.0
V 4.0.3
0.05
0.5
Fig. 8.11: Interference Signal
5
AMP 200N series
41 / 60