Ametek AMP 200N Manual For Operation page 49

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS
Test characteristics for SAE J1113-2 continuous interference
Method:
Coupling Method :
Test Regulation Method:
Disturbance Current limit : 1 A
Diagram Unit
Tolerances
Tolerances :
Additional Settings (Closed Loop with Scope only)
Level Change at next Testpoint:
Couplings
Coupling Type
Coupling Device
User Software settings
The user has to complete the test settings with his individual used coupling and measuring devices.
New coupling devices can be installed in the menu Setup/Coupling Device by the user.
A: Voltage Measuring Instrument:
Measurement Instrument
Scope
Instrument Value
B: Measuring Equipment (voltage probe):
A current transformer probe is used as current sensor connected to a measurement scope. This instrument
will measure the transferred ac voltage. The selected AC current instrument must be able to measure up to
250kHz
Equipment Type
Equipment
Manual for Operation
Conducted
Voltage Level (Closed Loop)
X-Axis :
kHz ; Log
Y-Axis:
Vpp ; linear
positive
+5%
negative
-5%
-3dB
Coupling Network
EM TEST CN 200N
Measuring AC voltage ( pk-pk)
Mandatory in SAE standard
Voltage RMS value
* Peak to Peak is critical for measuring with a scope
Voltage measurement
Voltage Probe
LeCroy HFP 2500
10:1
V 4.0.3
AMP 200N series
Current measurement
Current probe
Agilent N278A3
100mV/A
49 / 60