Immunity To Gloyd Gl Vi-7-2 - 20 (Dc); Dc Power Supply - Ametek AMP 200N Manual For Operation

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS
Immunity to GLoyd GL VI-7-2 – 20 (DC)
12.6

11.6.1. DC power supply

For loading this testfile the user has to open the file "GL VI-7-2 – 20 (DC).wim".
Press in the main Window "Immunity play"
1.
. Load the file "GL VI-7-2 – 20 (DC).wim". The program will open the test
2
Default test setup continuous interference
The software will show the picture of SAE J1113-2 with a very similar test setup.
Remark : The optional Shunt Capacitor is not part of the GLoyd test setup. In accordance with other standards
with similar test as in SAE J1113-2, MIL STD 461 CS 101 or RTCA DO-160D Section 18, it may be
necessary to use a shunt capacitor. The shunt capacitor is for shunt the dc source terminals for the
RF signal, if difficulty is encountered in obtaining sufficient test voltage.
GLoyd GL 2003 VI-Teil 7 Kapitel 2 Test Parameter requirements
Parameters for DC supply
Range
Frequency [kHz]
0.05 – 10
Range
Manual for Operation
Figure 28 Setup for GL VI-7-2 – 20
Figure 29 Signal voltage level for DC supply
Us [Veff]
Frequency step
3
1.5x10-3 dec/s (1% / 3s)
V 4.0.3
.
AMP 200N series
50 / 60

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Amp 200n1.1Amp 200n2Amp 200n1

Table of Contents