Ripple Immunity With Voltage And Current Measurement - Ametek AMP 200N Manual For Operation

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS
12.7

Ripple immunity with voltage and current measurement

NetWave
Signal Out
Source In
Framebus
Figure 30 Setup for ripple immunity with swapped voltage and current measurement channels
The above figure shows the setup for a closed-loop ripple immunity testing. The AMP 200N measurement
channels are used to measure the ripple current and voltage at the EUT terminals.
Recommended equipment:
Differential voltage probe
Active current probe
Shunt capacitor
Verification Resistor
Hints for the setup:
-
Use active voltage and current probes (to match the impedance of the AMP measuring channels).
Passive probes are not suitable.
-
Power the differential voltage probe with batteries. Using an external power supply may generate
interference that disturbs the ripple test.
-
AMP measurement channels can be swaped in case higher voltage measurement resolution is required.
In this case additional adapters from BNC to 4mm lab connectors (or vice versa).
-
It is recommended to first verify the setup and performance using a verification load instead of the EUT.
Manual for Operation
Computer
AC
DC
m
m
8
100
F
F
C-Box 100
AMP200Nx
Framebus
Signal Out
AMP 200N measuring input maximum allowed input voltages:
Voltage Measuring Input: max. 200 Vpp
Current Measuring Input: max. 10 Vpp
EM Test PVS 7 or Teseq MD 210
Keysight N2780B (2MHz, 500A), N2781B (10 MHz, 150A) or N2782B (50MHz,
30A) with external power supply
EM Test C-Box 10/100-1000
EM Test SVP CL 100 (100 Ohms)
CN 200N1
0. 5 R
Current
Probe
Power
BNC
4mm
to
to
Supply
BNC
4mm
V 4.0.3
AMP 200N series
+
EUT
_
Power
Supply
Verification
Load
52 / 60