AMETEK CTS
Test characteristics for CI 250 Transient interference
Signal
Adjust the
coupling factor
The used Solar transformer 6220-1A has a 2:1 turns ratio transformation at 1kHz. This ratio
will change with higher frequencies. At 100kHz the effective ratio is more like 16:1. EM Test
thinks the inter-winding capacitance results in a higher than anticipated drive voltage to
produce the +/-5V signal at the secondary.
Therefore the user has to adjust the transformer ratio of his transformer in the CN200N by
changing the Coupling Factor to approx. 10.7 in the setup shown in the upper figure. The
exact value must be measured during calibration.
Output
Settings as per
user definition
Select Channel
Channel 2 (DDS for AMP200N)
AMP 200N
2 Wire output ( High , Low)
Set End Voltage
As per user definition
Trigger
Settings as per
user definition
Coupling
CN 200N
Manual for Operation
V 4.0.3
AMP 200N series
46 / 60