Ametek AMP 200N Manual For Operation page 56

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS
Test characteristics
Method:
Coupling Method :
Test Regulation Method:
Diagram Unit
Tolerances
Tolerances :
Additional Settings:
Start at Level :
Level Change at next Testpoint:
If the first measurement is already too high the user can reduce the start Level ( example to -30dB)
User Software settings for adapt the hardware
The user has to complete the test settings with his individual used coupling and measuring devices. New
coupling devices can be installed in the menu Setup/Coupling Device by the user.
Coupling
Coupling Type
Coupling Device
Select the antenna you are using for the test.
Setup for add or modify new Antennas in the Library in the Setup Coupling Devices
 For more information refer to the Autowave software manual.
Manual for Operation
Default settings as per Standard
Radiated
Verify H-Field
X-Axis :
kHz ; Log
Y-Axis:
dBpT ; Log
+2.7% -0%
-20 dB
0dB
Antenna
Radiating Loop
V 4.0.3
AMP 200N series
56 / 60