Immunity To Ground Voltage Offset: Ci 250 - Ametek AMP 200N Manual For Operation

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS

12.4.5 Immunity to Ground Voltage Offset: CI 250

Components shall be immune from AC ground offset voltages. Requirements include both continuous and
transient disturbances.
Default test setup Continuous Disturbances
RI 250 Test Parameter requirements
Range
Frequency [kHz]
2.0 – 10
Range 1
Range 2
10 - 100
Manual for Operation
Figure 21 Setup for CI 250
Figure 22 Graph for CI 210
Us [Vp-p]
Frequency step [kHz]
1.0+1.14 log(f/10)
1.0
V 4.0.3
0.5
5
AMP 200N series
43 / 60