AMETEK CTS
10 Technical data AMP200N2
10.1
Amplifier output characteristics
Frequency range
Signal power
Power dissipation amplifier
Output voltage
Output current
DC current (for magnetic field tests)
Harmonic Distortion (THD)
Output Impedance
Current protection
Overtemperature protection
*DC into a 4 Ω load.
10.2
Signal generator output characteristics (built in)
Frequency range
Output voltage
Dc offset
10.3
Measurements (optional)
Measuring Unit AMP 200N
Frequency range
Accuracy
Current measurement
Voltage measurement
Magnetic field
10.4
General
Dimension
Weight
Supply voltage
Input power
Inrush current (power ON)
Fuses
Cooling
Manual for Operation
DC – 500kHz
1000 W* (nominal)
1000 W (maximum)
HIGH range:
±80Vp/
MED range:
±55Vp/
LOW range:
±25Vp/
HIGH range:
Max. 10.0A rms
MED range:
Max. 14.0A rms
LOW range:
Max. 18.0A rms
25 A
< 0.1%
30 mΩ @1 kHz
Short circuit protected
integrated into amplifier
DC, 10 – 500 kHz (sinusoidal)
±10 V
±10 V, programmable, to control external DC amplifier
Frequency-selective instrument for voltage, current and magnetic field
10 Hz – 250 kHz
(External oscilloscope must be used to measure beyond 250 kHz)
Better than 5%
External with current clamp
Range 100 mV/A:
Range 10 mV/A:
Max. 10 Vpp
17 mV – 70 V rms, max. 200 Vpp
Use voltage input for loop sensor measuring
19", 6 HU
(500mm x 449mm x 267mm)
40 kg
115 V +10/-2%, 50/60 Hz or (min 112 V)
230 V +10/-15%, 50/60 Hz
Max. 1000 W
60 A pk-pk
AMP 200N1:
115 V: 2 x 8 AT,
AMP 200N1.1
115 V: 16 AT,
Active cooling, air ventilation
V 4.0.3
1 mA – 30 A rms
10 mA – 300 A rms
230 V: 2 x 4 AT
230 V: 10 AT
AMP 200N series
26 / 60