Ametek AMP 200N Manual For Operation page 45

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS
Default test setup Transient Disturbances
Transient disturbances consist of a damped sinusoidal pulse with a resonant frequency of 100 kHz illustrated in
Figure 8.16. The pulse is applied using the delay sequence illustrated in Figure 8.17.
Delay times for the four sequences are listed in Table 18-2. Of Ford EMC-CS 2009.2 Standard.
CI 250 Test Parameter requirements
Note :
The Ford CS 2009.2 revision draft defines: The transient disturbances consist of a symmetrical damped sinusoidal
pulse with a resonant frequency of 100 kHz.
Manual for Operation
Figure 23 Setup for CI 250
Figure 24 Damped sinusoidal pulse
Figure 25 Damped sinusoidal with delay sequence d1...d20
V 4.0.3
AMP 200N series
45 / 60