Ametek AMP 200N Manual For Operation page 40

Low frequency signal source for supply simulation and magnetic fields
Table of Contents

Advertisement

AMETEK CTS
Test characteristics for RI 150
Default settings as per Standard
Method:
Coupling Method :
Test Regulation Method:
Diagram Unit
Tolerances
Tolerances :
Additional Settings:
System Impedance
Level Change at next Testpoint: -3dB
User Software settings
The user has to complete the test settings with his individual used coupling and measuring devices.
New coupling devices can be installed in the menu Setup/Coupling Device by the user.
A: Coupling
Coupling Type
Coupling Device
B: Current Measuring Instrument:
The current measuring is realized with a current probe connected to a measurement receiver. The AC
current instrument will measure the transferred ac voltage and must be able to measure up to 100kHz
Measurement Instrument
Scope
Instrument value
C: Measuring Equipment (current sensor):
The current sensor is a current transformer probe connected to a measurement receiver. This instrument
will measure the transferred ac voltage. The selected AC current instrument must be able to measure up to
100kHz
Equipment Type
Current Probe
Equipment
Agilent N2783A
Current Measuring on Channel 2:
Measuring Instrument (AC current)
1. Instrument
2. Instrument value
Measuring Equipment
3. Equipment Type
4. Equipment
For scope parameter settings please refer to the Autowave software manual.
Manual for Operation
Conducted
Current Level ( Closed Loop)
X-Axis :
kHz ; log
Y-Axis:
A pk-pk ; linear
+10% -0%
4.0Ω or used value
Test Fixture
Schwarzbeck TF130-150
R&S RSA TS-RI130
Measuring AC voltage
Mandatory in Ford EMC-CS 2009 standard
Voltage Peak to Peak
Transform the current into a rms ac voltage
Ratio 0.1V/A
Scope model
Voltage pk-pk
Current Probe
Agilent N2783A
V 4.0.3
AMP 200N series
40 / 60