8.6. Error Messages during CalKit Calibration and CubeScan Phantom Calibration
Error
Code
1
Insufficiency of saline solution
2
Transducer error
3
Plane motor error(1)
4
Plane motor error(2)
5
Reserved
6
Reserved
7
Error in the Analog Board
8
If there' s no reflection signal from
target.
9
Angle Motor Error(1)
10
Angle Motor Error(2)
11
Distance error from transducer to
target
12
Too low echo signal
13
Too high echo signal
14
Reserved
15
Measured volume value is different
reference value.
16
Scan data transmission error
17
Error in detection of bladder phantom
sides.
18
Reserved
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Description
Remark
Check the solution level.
Check the transducer
Check probe connection
Check the plane motor
Check probe connection
Check the plane motor
Check probe connection
Check the Analog Board
Probe is not in the CalKit.
Check the angle motor
Check probe connection
Check the angle motor
Check probe connection
Frame – out of position
Confirm that the air bubble is
absent.
Check the Analog Board
Check the Analog Board
(Phantom calibration only)
Check the Control Board
(Phantom calibration only)
(Phantom calibration only)