Tcg2 Test - Mcube CubeScan BioCon-500 Service Manual

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500

5.1.3 TCG2 Test

① Turn on the analog board.
② Set up Oscilloscope as below and connect the probe to TP11.
③ Press the scan button of BioCon-500
④ Measure the rising time of the voltage measured with Oscilloscope from 0.5V
to 1.5V.
- The signal is measured in the following figure.
- At this time, the value during the rising time must be measured as below.
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Item
Axis of Voltage
Axis of Time
Trigger Level
Trigger Method
(The Figure of TCG2 Measurement Signal)
Item
Rising Time
Setup Value
0.5 [V]
40 [us]
+500 [mV]
Single Shot
Measured Value
156 ~ 172 [us]
39