Service Manual for BioCon-500
Ultrasonic Pulse
5.1.4
① Turn on the analog board.
② Set up the Oscilloscope as below and connect the probe to TP6.
③ Press the scan button of BioCon-500
④ Measure the voltage shape measured by Oscilloscope
-The signal is measured in the following figure.
- The peak-to-peak voltage at TP6 must be measured in the following scope.
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Test
Item
Axis of Voltage
Axis of Time
Trigger level
Trigger Method
(Figure of the Ultrasonic Pulse Signal)
Item
Vpp @TP6
Setup Value
20 [V]
400 [ns]
+20 [V]
Single Shot
Measured Value
60 ~ 90 [V]