Service Manual for BioCon-500

5. Signal Test

5.1 Test for the Analog Board
To check and test the signal of the analog board, following measurement devices
are required.
- Digital multi-meter
- Digital oscilloscope
Test the analog board by checking the signal with above measurement devices in
the following order.

5.1.1 Power Test

① Turn on the analog board.
② Test the power for below location by using multi-meter.
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Location
TP1 & '+' of C40
TP1& '-' of C48
TP1& TP8
Measured Value
+5V (+/- 5%)
-5V (+/- 5%)
+7V (+/- 5%)
37