Service Manual for BioCon-500
8. Calibration Procedures
8.1 Purpose
This procedure is prepared for guidance on the calibration of the BioCon-500. After fully
understanding of this procedure, do the calibration.
8.2 Scope
This procedure covers preparations & procedures for needed to calibrate the BioCon-500
using CalKit, CubeScan Phantom, or Dansk Phantom.
8.3 Supported Firmware Version
l
Version 3.10pt, Version 3.10pto, Version 3.10apt
l
Version 2.61pt, Version 2.61pto, Version 2.61apt
8.4 General
-
CalKit calibration should be done in a normal mode (not in a maintenance mode).
- CubeScan phantom calibration and Dansk phantom calibration should be done in a
maintenance mode. There is no way to do CubeScan phantom calibration and Dansk
phantom calibration in a normal mode.
- There is no limit in using even though the calibration time is elapsed since last
calibration currently. But this policy can be changed without prior notice.
- Do not mix up the calibration menu and the tool used in a calibration.
- Normal mode vs. Maintenance mode. You can switch two modes using the combined
key ' ENTER+LEFT+PRINT'. In a maintenance mode the message "MAINTENANCE
MODE" is displayed in the Top Screen.
- Calibration necessity
Item
1
When you upgraded BioCon-500' s program over
Version 3.10
2
After replacing Analog Board
3
After replacing Digital Board
4
After replacing Probe
5
Periodical check (Once a year )