Calibration Procedures; Purpose; Scope; Supported Firmware Version - Mcube CubeScan BioCon-500 Service Manual

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500

8. Calibration Procedures

8.1 Purpose

This procedure is prepared for guidance on the calibration of the BioCon-500. After fully
understanding of this procedure, do the calibration.

8.2 Scope

This procedure covers preparations & procedures for needed to calibrate the BioCon-500
using CalKit, CubeScan Phantom, or Dansk Phantom.

8.3 Supported Firmware Version

l
Version 3.10pt, Version 3.10pto, Version 3.10apt
l
Version 2.61pt, Version 2.61pto, Version 2.61apt

8.4 General

-
CalKit calibration should be done in a normal mode (not in a maintenance mode).
- CubeScan phantom calibration and Dansk phantom calibration should be done in a
maintenance mode. There is no way to do CubeScan phantom calibration and Dansk
phantom calibration in a normal mode.
- There is no limit in using even though the calibration time is elapsed since last
calibration currently. But this policy can be changed without prior notice.
- Do not mix up the calibration menu and the tool used in a calibration.
- Normal mode vs. Maintenance mode. You can switch two modes using the combined
key ' ENTER+LEFT+PRINT'. In a maintenance mode the message "MAINTENANCE
MODE" is displayed in the Top Screen.
- Calibration necessity
Item
1
When you upgraded BioCon-500' s program over
Version 3.10
2
After replacing Analog Board
3
After replacing Digital Board
4
After replacing Probe
5
Periodical check (Once a year )
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
When
Remark
Necessary
Necessary
Necessary
Necessary
Recommended
54