5.9

Calibrating the BioCon-500 Using the Calibration Kit

5.9.1
Preparing the calibration kit
1) Open the cap of the calibration kit and pour 0.9 % saline solution up to the
fill-mark of the calibration kit.
2) Close the cap.
3) Align the probe scan button with the arrow mark on the calibration kit, then
insert the probe into the holder ensuring it is secure.
IMPORTANT:
5.9.2
Calibrating the BioCon-500 with the calibration kit
1) Connect the probe which is positioned into the probe holder of the calibration
kit to the console
2) Turn on the console
Screen Display in the BioCon-500
Purge all air bubbles form the system.
You can also use distilled or DI water instead of the saline.
Purge the system of air bubbles in this case as well.
57
Description
The Top Screen.
Press UP to go to setup menu.
Setup Menu
Go to "Calibration" setup using UP or
DOWN.