Mcube CubeScan BioCon-500 Service Manual page 65

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500
8.5.2.2 Calibrating the BioCon-500 with the CubeScan Phantom
1) Connect the probe which is put into the probe holder of the CubeScan
phantom.
2) Turn on the BioCon-500
3) Check if the BioCon-500 is in a maintenance mode.
Screen Display in the BioCon-500
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
-
Description
The Top Screen.
Press UP key to go to setup menu.
Check that the system is in a
maintenance mode.
Press "LEFT+RIGHT" key.
Maintenance Top Screen
Press ENTER key.
Maintenance Setup Menu
1. Go to "Calibration" setup using UP or
DOWN key.
2. Press ENTER key.
3. Display the "Start" setup value using
LEFT key or RIGHT key.
Maintenance Calibration Top screen
Select CubeScan Phantom Calibration
using UP or DOWN key and press
ENTER key.
60