The system variables of Maintenance mode have the following meanings.
Item
Cal. Value
Store Mode
System Reset
Probe Offset
Print Density
Outline Mode
Exit
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Change the threshold value. Value range is from 5 to 20.
This value is changed during calibration process.
Appoints which one to store from the current session. In case of
current, the last scanned value is stored. And in case of maximum, the
image data with maximum volume is stored.
Reset values of system variables to default.
The default values are as below.
Menu
Cal. Value
Store Mode
Maintenance
Probe Offset
Print Density
Outline Mode
Print Option
Flash Store
Setup
Scan Result
Auto Power
Prescan Enable
Set the initial position of the transducer in the probe. Range is from 3 to
12.
This value is set during calibration process.
Adjusts the density of the print out. The adjustable value ranges from 3
to 10.
On / Off control
To control whether the edge information of the bladder is displayed or
not on the screen with ultrasound image.
On : The edge information will be displayed.
Returns to the previous screen in Maintenance menu.
Contents
Item
Reset Value
10
Maximum
9
7
On
Raw Image
On
B-Mode
5 minute
On
53