JP3 is a connector linked to the ultrasonic probe. And it consists of signals for a
driving of a stepping motor built in the ultrasonic probe, a signal for transmission and
reception of the electric signals with ultrasonic oscillator, and a signal for reception of
the input from the switch built in the ultrasonic probe. These signals are defined as
below.
TO:
Pin No.
1
2
3
SCAN_SW & DATA
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
*. JP7, JP4, JP5: Refer to JP1, JP2, and JP3 of Control Board, respectively.
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Name
Shield ground
CHECK_CON2
-
ANGLE1
CHECK_CON2
-
ANGLE2
-
ANGLE3
ANGLE4
-
PLANE1
-
PLANE2
PLANE3
PLANE4
SIGNAL
-
SIG_GND
JP3
Ultrasonic Probe
I/O
Description
Power
Cable shield
in
Check Connector 2(Orange)
In
Scan switch input from probe (Grey)
-
Not Connected
Out
Step motor control signal (Red)
in
Check Connector 2 (Violet)
-
Not Connected
Out
Step motor control signal (Blue)
-
Not Connected
Out
Step motor control signal (White)
Out
Step motor control signal (Black)
-
Not Connected
Out
Step motor control signal (Brown)
-
Not Connected
Out
Step motor control signal (Green)
Out
Step motor control signal (Pink)
Out
Step motor control signal (Yellow)
I/O
Signal line (Coaxial +)
-
Not Connected
Power
Signal ground (Coaxial Shield)
12