Mcube CubeScan BioCon-500 Service Manual page 62

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500
IMPORTANT:
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
l
The BioCon-500 returns to the top screen when calibration
process is terminated or stopped.
l
To stop the calibration menu, press the down key (only
effective in the first and second path of the calibration
Calibration top screen
To start calibration, press ENTER key.
Screen during calibration.
To stop calibration, press DOWN key
in the first path or second path.
In the third path and fourth path, a
user cannot stop the calibration.
When there is any error during
calibration, the system displays the
following message in the bottom part
of the LCD screen.
"Calibration has failed!!!"
"ERROR: First Path, Not Enough
Water (EC1)".
In the error message, "First Path"
means the calibration steps when the
system detected any error during
calibration.
"Not Enough Water" means additive
information about the error state. This
additive information is displayed for
only EC1 error.
"EC1" means the error code during
calibration.
The screen when the calibration is
done successfully. The date and time
when
the
calibration
successfully is displayed in the bottom
part of the LCD screen.
is
done
57