Mcube CubeScan BioCon-500 Service Manual page 52

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500
6.4 The green LED is not turned on even when the power adapter is
connected.
No.
Is green LED not turned on even when the
1
system is OFF?
Is the 4 pin connector connecting the Key
2
Board and the Charge Board connected?
3
Green LED is on when the system is OFF,
4
but it is turned off when the system is ON.
6.5 NO SCANHEAD message is displayed during scanning.
No.
Is the ultrasonic probe connected well to
1
the Console?
Is open between the 3
2
probe connector?
The message is displayed when the 3
3
th
and 6
pin of Probe connector is short
6.6 NO PAPER message is displayed during printing.
No.
1
Does printer module have thermal paper?
Does message still appears even when
2
there is paper?
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Check point
Check point
rd
th
and 6
pin of the
Check point
Description
Yes ->Refer to item 2.
No -> Refer to item 4
Check the connection status and if not
connected,
connect
connector.
Replace the Charge Board.
No problem.
Because of charging control in Control
board sometime Green LED is off
when system is on.
Description
If not, NO SCANHEAD message is
displayed when the scan button is
pushed.
Test again after replacing with other
ultrasonic probe.
rd
Request service from the headquarter
Description
If not, insert thermal paper.
Try again after replacing the control
board.
the
4
pin
47