6.7 No image is displayed after scanning.
No.
1
2
3
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Check point
Description
In this case, the mode showing the
scan result is set as CONTOUR.
Normal Image
Scanned image from open object (in
air)
Image
when
the
connection is open.
Check the resistance from the
ultrasonic probe connector to the
th
th
18
and 20
pin
Error in transducer connection.
Retry after replacing the probe
Check if ' SCAN RESULT' is set as
' CONTOUR' .
Yes, retry after setting as ' B-MODE' .
No,
request
service
headquarter.
transducer
from
the
48