SERVICE TRAINING
Mcube Technology Co., Ltd.
www.mcubetech.co.kr

Advertisement

loading

Summary of Contents for Mcube CubeScan BioCon-500

  • Page 1 SERVICE TRAINING Mcube Technology Co., Ltd. www.mcubetech.co.kr...
  • Page 2 BioCon- - - - 500 Service Training BioCon 500 Service Training BioCon BioCon 500 Service Training 500 Service Training Contents 1. General System 2. System Configuration 3. System Setup 4. Module Test 5. Troubleshooting www.mcubetech.co.kr...
  • Page 3 1. General General General General PRINCIPLES (1) i=n-1 ∑ Gain Lateral Position 125mL ultrasonic reflections 3-dimensional abstract calculate scanning bladder wall bladder volume www.mcubetech.co.kr...
  • Page 4 1. General General General General PRINCIPLES (2) ▣ ▣ ▣ ▣ Measuring ultrasonic reflections within a patient's body - differentiates the urinary bladder from the surrounding tissues. ▣ ▣ ▣ ▣ Measuring ultrasonic reflections on multiple planes - 3-dimensional scanning ▣...
  • Page 5 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration System block diagram Connection diagram Device disassembly www.mcubetech.co.kr...
  • Page 6 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration System block diagram www.mcubetech.co.kr...
  • Page 7 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Connection diagram Charge JP101 Analog Board Board Ultrasonic Adapter Probe Battery Connection Board Battery Module Board Printer Module Control Board Key Board LCD Module www.mcubetech.co.kr...
  • Page 8 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Components of the Device LCD Module / Printer Module / LCD Module / Printer Module / LCD Module / Printer Module / LCD Module / Printer Module / Upper Case Upper Case Upper Case...
  • Page 9 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Console www.mcubetech.co.kr...
  • Page 10 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Upper Case www.mcubetech.co.kr -10-...
  • Page 11 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Lower Case www.mcubetech.co.kr -11-...
  • Page 12 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Ultrasonic Probe Module(1) www.mcubetech.co.kr -12-...
  • Page 13 2. System Configuration 2. System Configuration 2. System Configuration 2. System Configuration Ultrasonic Probe Module(2) 1 1 1 1 3 3 3 3 5 5 5 5 7 7 7 7 9 9 9 9 white white white white Coaxial Coaxial Coaxial Coaxial- - - -...
  • Page 14 3. System Setup 3. System Setup 3. System Setup 3. System Setup Firmware Upgrade Set the Maintenance Menu Set the Setup Menu Calibration www.mcubetech.co.kr -14-...
  • Page 15 3. System Setup 3. System Setup 3. System Setup 3. System Setup Firmware Upgrade Step Activity Install “CS_Maintenance CS_Maintenance” to a PC. CS_Maintenance CS_Maintenance Execute “CS_Maintenance CS_Maintenance CS_Maintenance CS_Maintenance”. Connect the device to the PC by using a USB cable. Turn on the device.
  • Page 16 3. System Setup 3. System Setup 3. System Setup 3. System Setup Maintenance Menu Item Available Value Reset Value Cal. Value 5~20 Store Mode Current / Maximum Maximum Return / Reset System Reset Return (The scan data stored is deleted) Probe Offset 4 ~ 20 Print Density...
  • Page 17 3. System Setup 3. System Setup 3. System Setup 3. System Setup Setup Menu Item Available Value Reset Value Clinic Name Set Date / Time Value Only / Raw Image / Walls / Value Only / Raw Image / Walls / Value Only / Raw Image / Walls / Value Only / Raw Image / Walls / Print Option...
  • Page 18 3. System Setup 3. System Setup 3. System Setup 3. System Setup Calibration 1. Check the error in the ultrasound transducer. 2. Verify the start position of the plane motor. 3. Verify the direction of rotation of the plane motor. 4.
  • Page 19 3. System Setup 3. System Setup 3. System Setup 3. System Setup Required material • BioCon-500 Unit (Including Probe) • Calibration Kit • Saline Solution: About 2.5 liter www.mcubetech.co.kr -19-...
  • Page 20 3. System Setup 3. System Setup 3. System Setup 3. System Setup When we should perform calibration? Item Item Item Item When When When When Remark Remark Remark Remark When you upgraded BioCon-500’s program over Necessary Version 3.10 After replacing Analog Board Necessary After replacing Digital Board Necessary...
  • Page 21 3. System Setup 3. System Setup 3. System Setup 3. System Setup Preparing the CalKit..Step Figure Description Open the cap of the CalKit and pour saline solution up to the fill- mark of the CalKit. Confirm that the air bubble is absent.
  • Page 22 3. System Setup 3. System Setup 3. System Setup 3. System Setup Preparing the CalKit..Step Figure Description Close the cap of the CalKit. www.mcubetech.co.kr -22-...
  • Page 23 3. System Setup 3. System Setup 3. System Setup 3. System Setup Preparing the CalKit..Step Figure Description Align the probe scan button with the arrow mark of the Calkit, and put the probe into the probe holder firmly.
  • Page 24 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description The Top Screen. Connect probe and then turn on the BioCon-500. Press UP key to go to setup menu. www.mcubetech.co.kr -24-...
  • Page 25 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description Setup Menu Go to “Calibration” setup using UP or DOWN key. www.mcubetech.co.kr -25-...
  • Page 26 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description Press ENTER key. www.mcubetech.co.kr -26-...
  • Page 27 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description Display the “Start” setup value using LEFT key or RIGHT key www.mcubetech.co.kr -27-...
  • Page 28 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description Calibration top screen To start calibration, press ENTER key. www.mcubetech.co.kr -28-...
  • Page 29 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description Screen during calibration. To stop calibration, press DOWN key in the first path or second path. In the third path and fourth path, a user cannot stop the calibration www.mcubetech.co.kr -29-...
  • Page 30 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description When there is any error during calibration, the system displays the following message in the bottom part of the LCD screen. *Common points to be checked in case *Common points to be checked in case *Common points to be checked in case *Common points to be checked in case...
  • Page 31 3. System Setup 3. System Setup 3. System Setup 3. System Setup Processing calibration Step Screen Display Description The screen when the calibration is done successfully. The date and time when the calibration is done successfully is displayed in the bottom part of the LCD screen.
  • Page 32 3. System Setup 3. System Setup 3. System Setup 3. System Setup Calibration Error Code Error Error Description Description Remark Remark Error Error Description Description Remark Remark Code Code Code Code Insufficiency of saline solution Check the water level. Transducer error Check the transducer Check probe connection Variation in the plane motor start...
  • Page 33 3. System Setup 3. System Setup 3. System Setup 3. System Setup Calibration Error Code Error Error Description Description Remark Remark Error Error Description Description Remark Remark Code Code Code Code Error in the Analog Board Check the Analog Board If there’s no reflection signal from Probe is not in the CalKit.
  • Page 34 3. System Setup 3. System Setup 3. System Setup 3. System Setup Calibration Error Code Error Error Description Description Remark Remark Error Error Description Description Remark Remark Code Code Code Code Distance error from transducer to Frame breakaway target Confirm that the air bubble is Too low echo signal absent.
  • Page 35 3. System Setup 3. System Setup 3. System Setup 3. System Setup Calibration Error Code Error Error Description Description Remark Remark Error Error Description Description Remark Remark Code Code Code Code Measured volume value is (Phantom calibration) different reference value. Scan data transmission error Check the Control Board Check the Analog Board...
  • Page 36 4. Module Test 4. Module Test 4. Module Test 4. Module Test Analog Board Ultrasonic Probe Control Board www.mcubetech.co.kr -36-...
  • Page 37 4. Module Test 4. Module Test 4. Module Test 4. Module Test Analog Board Item Location Value Power TP1(GND) & '+' of L9 +5V (+/- 5%) Power TP1(GND) & '-' of L3 -4.8V (+/- 5%) Power TP1(GND) & TP8 +7.7V (+/- 5%) TCG1 Rising Time TP1(GND) &...
  • Page 38 4. Module Test 4. Module Test 4. Module Test 4. Module Test Ultrasonic Probe Probe Cable Pin No. Measured Value Pin 5 & 8 Pin 10 & 11 15(+/-)1 (ohm) Pin 13 & 15 Pin 16 & 17 Pin 1 & 6 Short (normal) /Open (abnormal) Pin 1 &...
  • Page 39 4. Module Test 4. Module Test 4. Module Test 4. Module Test Control Board Check point Check operation of the LCD. Check operation of the Printer. Check operation of the Key Board. Check operation of the Ultrasonic Probe. Check operation of the USB. www.mcubetech.co.kr -39-...
  • Page 40 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting System is not turned on Check point Description Is POWER key pressed for more than 1 The system operates when pressing the second? POWER key for more than 1 second. Voltage of battery module. Above +7V Is battery module connected? If not, connect the module.
  • Page 41 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting System is not turned off Check point Description The system operates when Is POWER key pressed more pressing POWER key for more than 1 second? than 1 second. Turn the power off by Is the system not turned off removing the battery module even when pressing the Power...
  • Page 42 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting System turn off while in use Check point Description When auto power-off function is active, system Is auto power-off on in the will be turned off if no key setup menu? input is made during the fixed time.
  • Page 43 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Initial screen is not display after turn-on Check point Description Check if the connector for backlight Is backlight on in LCD? of LCD is connected. Is the FPC connector to the LCD module Check if the connector is connected well connected? properly.
  • Page 44 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Green LED is not turned on with adapter being connected Check point Description Is green LED not turned on even Yes ->Refer to item 2. when the system is OFF? No -> Refer to item 4 Is the 4 pin connector connecting the Check the connection status and if Key Board and the Charge Board...
  • Page 45 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting NO SCANHEAD Error Message Check point Description If not, NO SCANHEAD Is the ultrasonic probe message is displayed when connected to the main body? the scan button is pushed. Is open between the the 1 Test again after replacing pin of the probe connector? with other ultrasonic probe.
  • Page 46 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting NO PAPER Error Message Check point Description Does printer module have If not, insert thermal paper. thermal paper? Does message still appears Try again after replacing the even when there is paper? control board.
  • Page 47 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Image after SCAN(1) • Image showing the scan result is set as CONTOUR • Normal Image • Scanned image from open object (in air). www.mcubetech.co.kr -47-...
  • Page 48 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Image after SCAN(2) • Image ,when the transducer connection is open. • Check using the Calibration Kit. • Error in transducer connection. • Retry after replacing the probe www.mcubetech.co.kr -48-...
  • Page 49 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Image after SCAN(3) • Check if ‘SCAN RESULT’ is set as ‘CONTOUR’. • Yes, retry after setting as ‘B-MODE’. • No, request service from the headquarter. www.mcubetech.co.kr -49-...
  • Page 50 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Error in Motor Rotation Check point Description Check error in motor. Measure the resistance of each case in probe Connector Normal resistance value 1) Pin number 5 and 8 15 +/- 1 (Ohms) 2) Pin number 10 and 11 3) Pin number 13 and 15 4) Pin number 16 and 17...
  • Page 51 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Printing Error (No feeding) Check point Description No paper. Thermal Paper Properly insert. Yes, goto 3 Motor sound? No, check the printer connector Not solved? Replace control board. *Paper Spec. : 57mm(width), ø30mm *Paper Spec.
  • Page 52 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting 5. Troubleshooting Battery Management 1) After full charging (for battery module with full capacity) 1) After full charging (for battery module with full capacity) 1) After full charging (for battery module with full capacity) 1) After full charging (for battery module with full capacity) scan : 4 hours(one scan per 15sec) scan : 4 hours(one scan per 15sec)