Assembling Of The Lower Case - Mcube CubeScan BioCon-500 Service Manual

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500

3.2.2 Assembling of the Lower Case

The assembling of the Lower Case takes the following steps. And the steps are
reversed for the disassembling.
Step
Attach the ⑦Spacers to the ①Lower Case with 10mm from the
1
①Lower Case edge.
Attach the ⑧Handle by using the ⑩M3 Washer Bolts.
2
Attach the ②Control Board, the ③Analog Board, the ④Charge
Board, and the ⑤Battery Board to the ①Lower Case by using the
3
⑩M3 Washer Bolts.
At this time, the upper right hole of the ②Control Board and the
lower right hole of the ③Analog Board must be attached by the
4
⑨Supporter.
Connect the ⑥Connection Board into the ②Control Board and
the ③Analog Board in the correct direction, and attach them by
5
using ⑩M3 Washer Bolts.
Connect the ⑥Charge Board and the ③Analog Board by using
6
the ⑪Power Cables.
Connect the ⑥Charge Board and the ⑤Battery Board by using
7
the ⑫Battery Cable.
The picture of the assembled lower case is as below.
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Description
(The Picture of the Lower Case Assembled)
Remark
⑦ X 4EA
⑩ X 2EA
⑩ X 16EA
⑨ X 2EA
⑩ X 2EA
⑪ X 1SET
⑫ X 1EA
29