Mcube CubeScan BioCon-500 Service Manual page 54

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500
6.8 The system is turned off while in use.
No.
Is auto power-off on in the setup menu of
1
Maintenance mode?
Is the system turned off with the message
2
of "BATTERY LOW, SYSTEM WILL BE
TURNED OFF"?
6.9 Error in motor rotation during scan operation
No.
Check error in motor.
Measure the resistance of each case in
probe Connector
1
1)
Pin number 5 and 8
2)
Pin number 10 and 11
3)
Pin number 13 and 15
4)
Pin number 16 and 17
2
In case of normal resistance
6.10 Error in Printing (No feeding)
No.
1
Thermal Paper
2
Motor sound?
3
Not solved?
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Check point
Check point
Check point
Description
When auto power-off function is
active, system will be turned off if no
key input is made during the fixed
time.
Turn off the system to protect the
battery
module
if
the
insufficient. Try again after charging.
Description
Normal resistance value
15 +/- 1 (Ohms)
Test again after replacing with other
analog board
Description
No paper.
Properly inserted.
Yes, go to 3
No, check the printer connector
Replace control board.
power
is
49

Hide quick links: