Service Manual for BioCon-500

8.5.2 CubeScan Phantom Calibration

8.5.2.1 Preparing the CubeScan phantom
No.
Figure
1
2
3
4
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Place the CubeScan phantom on a flat surface.
And open the cover of the phantom.
Drop about 5ml water to the center surface of
the phantom.
Place the holder on the top of the Cubescan
phantom.
Check if the holder is in a stable and flat
position.
Put the probe into the probe holder.
Description
59