Procedures; Calkit Calibration - Mcube CubeScan BioCon-500 Service Manual

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500

8.5 Procedures

8.5.1 CalKit calibration

8.5.1.1 Preparing the CalKit
No.
Figure
1
2
3
8.5.1.2 Calibrating the BioCon-500 with the calibration kit (CalKit)
1) Connect the probe which is put into the probe holder of the calibration kit.
2) Turn on the BioCon-500
3) Check if the BioCon-500 is in a normal mode.
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
Open the cap of the calibration kit and pour
saline solution up to the fill-mark of the
calibration kit.
Confirm that the air bubble is absent.
You can also the pure water on behalf of the
saline solution. But in this case, do calibration
after confirming that all the air bubbles went
away.
Close the cap of the calibration kit.
Align the probe scan button with the arrow mark
of the calibration kit, and put the probe into the
probe holder firmly.
Description
55