Mcube CubeScan BioCon-500 Service Manual page 66

Hide thumbs Also See for CubeScan BioCon-500:
Table of Contents

Advertisement

Service Manual for BioCon-500
IMPORTANT:
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
l
The BioCon-500 returns to the top screen when calibration
process is terminated or stopped.
l
To stop the calibration menu, press the down key (only
effective in the first and second path of the calibration
process) more than 5 seconds or turn of the system by
pressing the power key.
To start calibration, press ENTER key.
Screen during calibration.
To stop calibration, press DOWN key in
the first path or second path.
In the third path and fourth path, a user
cannot stop the calibration.
When
there
is
calibration, the system displays the
following message in the bottom part of
the LCD screen.
"Calibration has failed!!!"
"ERROR: First Path(EC3)".
In the error message, "First Path" means
the calibration steps when the system
detected any error during calibration.
"EC3" means the error code during
calibration.
The screen when the calibration is done
successfully. The date and time when the
calibration
is
done
displayed in the bottom part of the LCD
screen.
any
error
during
successfully
is
61