Service Manual for BioCon-500

4.2 Ultrasonic Probe Cable

The cable between the Console and the Ultrasonic Probe consists of ultrasonic, motor
control, and user input signals as the following structure. In the figure below, D is a
terminal connected to the Console, while B and C are the terminals located within the
Probe.
The signal configuration at the Ultrasonic Probe Cable is as below.
D
B
(Pin No.)
(Pin No.)
1
2
3
4
5
1
6
7
8
2
9
10
3
11
4
12
13
9
14
15
10
16
7
17
8
18
6
19
20
5
Mcube Technology Co., Ltd.
www.mcubetech.co.kr
TM
(Configuration of the Ultrasonic Probe Cable)
C
Color
(Pin No.)
orange
grey
-
red
violet
-
blue
-
white
black
-
brown
-
green
pink
yellow
-
(Pin Configuration of the Ultrasonic Probe Cable)
Description
Cable shield
Check Connector 2
Scan Switch input from probe
Not Connected
Control signal for step motor
Check Connector 1
Not Connected
Control signal for step motor
Not Connected
Control signal for step motor
Control signal for step motor
Not Connected
Control signal for step motor
Not Connected
Control signal for step motor
Control signal for step motor
Control signal for step motor
Signal line (Coaxial +)
Not Connected
Signal ground (Coaxial shield)
35