8.3 Carry out crockmeter tests

Never touch any moving parts during operation.
The side panels must always be mounted during operation.
Do not lean over the unit during operation.
Risk of injury: Never hold the sample by hand during operation. It has to be
fixed with an appropriate means, e.g. double sided adhesive tape.
Place the device on a solid, even surface.
Level the apparatus using the included spirit level. To do so, place the spirit
level on the sample tray and adjust the levelling feet (9) until the Proceq ZAA
2600 is levelled perfectly. The sample tray must be firmly in place.
Fold up the tool holder(4).
Open the screws of the sample frames (6) with a quarter-turn and remove the
sample frames (6).
Place the sample.
Place the sample frames (6) in the unit again and while pressing them down
fasten the screws with a quarter-turn.
Fold down the tool holder (4).
Page 41