Converting From Scrub Tester Into Basic Unit - Proceq ZAA 2600.SW Instruction Manual

Automatic scrub tester
Table of Contents

Advertisement

9

Converting from scrub tester into basic unit

Thanks to its modular construction, the automatic universal unit can be equipped
with different kits for carrying out applications, washability, crocking and scratch
resistance tests as well as recording of the drying time.
Before mounting an optional modification set, the device has to be converted
to basic unit.
Always unplug the power cable before assembling and/or converting the unit.
Fold up the tool holder (4) and open the screws of the sample frames (6) with a
quarter-turn.
Remove the sample frames (6) and afterwards the sample tray (3) together
with the drain hose (16).
Push the start stop (1) and the end stop (2) to the centre of the device and
remove them afterwards.
Loosen the fixing screws of the side panels (18) and remove the side panels
(17).
Loosen the Allen screws of the height adjustment and remove the tool holder
(4).
Loosen the screws of the fixing bar (20) and remove the fixing bars.
Screw in the four rubber feet delivered in place of the fixing bars.
Remount the side panels (17), the one with slit on the right side, the other one
on the left side.
Remount the previously removed screws of the start- (1) and end stop (2).
Check if the four rubber feet are mounted.
Basic unit
Page 44