Function "Measured Level" (0A6); Function "Detection Window" (0A7) (From Software 01.04.00) - Pepperl+Fuchs Pulscon LTC Description Of Instrument Functions

Guided level radar with hart/4 ma ... 20 ma and profibus pa
Table of Contents

Advertisement

Pulscon LTC
Function group "diagnostics" (0A)
11.7
+
E
Display of measured level in the selected "distance unit" (0C5).
11.8
+
E
Is used to switch the detection window on and off and to reset an existing detection
window. If this function is switched on, a window is defined surrounding the current level
echo (typical width: 1 m ... 2.5 m; depending on the application parameters). The
window always moves together with a rising a falling echo.
Echos beyond the limits of the window are ignored for a certain time.
Selection:
off
on
reset
After selection of this option, the current window is reset, the level echo is looked for
in the complete measuring range and a new window is defined surrounding the
current level echo.

Function "measured level" (0A6)

20 mA
100 %
measured dist.(0A5)
measured level (0A6)
4 mA
0 %

Function "detection window" (0A7) (from Software 01.04.00)

63