Function "Linearisation" (041) - Pepperl+Fuchs Pulscon LTC Description Of Instrument Functions

Guided level radar with hart/4 ma ... 20 ma and profibus pa
Table of Contents

Advertisement

Pulscon LTC
Function group "linearisation" (04)
+
E
32
6.2

Function "linearisation" (041)

Linearisation defines the ratio of level to container volume or product weight and allows
a measurement in customer units, e. g. metres, hectolitres etc. The measured value in
(000) is then displayed in the selected unit.
This function is used to select the linearisation modes.
Selection:
linear
horizontal cyl
manual
semi-automatic
table on
clear table
linear
The tank is linear e. g. a cylindrical vertical tank. You can measure in customer units by
entering a maximum volume/weight.
You can select the "customer unit" (042). Define the volume value corresponding to
the calibration in "max. scale" (046). This value corresponds to an output of
100 % (= 20 mA for HART).
20 mA
4 mA
100 %
0 %
3
3
kg, m
, ft
, ... = customer unit (42)
max.scale
(046)