Function Group "Linearisation" (04); Function "Level/Ullage" (040) - Pepperl+Fuchs Pulscon LTC Description Of Instrument Functions

Guided level radar with hart/4 ma ... 20 ma and profibus pa
Table of Contents

Advertisement

Pulscon LTC

Function group "linearisation" (04)

6
+
E
6.1
+
E
Selection:
level CU
level DU
ullage CU
ullage DU
level CU
Level in customer units. The measured value can be linearised. The
"linearisation" (041) default value is set to a linear 0 % ... 100 %.
level DU
Level in the selected "distance unit" (0C5).
ullage CU
Ullage in customer units. The value can be linearised. The "linearisation" (041) default
value is set to a linear 0 % ... 100 %.
ullage DU
Ullage in the selected "distance unit" (0C5).
Note!
!
Reference point for the ullage is "full calibr." (006) (= span).
Function group "linearisation" (04)

Function "level/ullage" (040)

20 mA
100 %
4 mA
0 %
Ullage
Level