Function "Probe Length" (033); Function "Determine Length" (034) - Pepperl+Fuchs Pulscon LTC Description Of Instrument Functions

Guided level radar with hart/4 ma ... 20 ma and profibus pa
Table of Contents

Advertisement

Pulscon LTC
Function group "length adjustment" (03)
+
E
+
E
30
5.4

Function "probe length" (033)

Use this function to enter the probe length.
5.5

Function "determine length" (034)

Use this function to determine the probe length automatically. Due to the mounting
conditions, the automatically determined probe length may be larger than the actual
probe (typically 20 mm ... 30 mm longer). This has no influence on the measuring
accuracy. When entering the empty value for a linerisation, please use the "empty
calibration" instead of the automatically determined probe length.
Selection:
length ok
too short
too long
After selecting "length too short" or "length too long" the calculation of the new probe
length takes up to approx. 10 s.